3.9 Sensor installation

3.9.1 Sensor positioning information

The sensor must be installed in a socket or flow chamber that allows contact with the sample
fluid to be analyzed.
The sensor and measuring instrument are connected by a cable. The standard sensor cable
lengths are 3, 5, 10, 15 and 20 meters.
Ensure that the sensor will be mounted:
perpendicular to the pipe
horizontal
on a horizontal pipe section (or on flow-ascending vertical pipe)
minimum of 15 meters away from the pump's discharge side
in a place where the sample flow is stable and rapid, and as far as possible from:
Note: There may be situations where not all the above conditions can be met. If this is the case, or you
have any concerns, please consult your Hach representative to appraise the situation and define the best
applicable solution.

3.9.2 Recommended sample flow rate

For optimal response time, the recommended sample flow rate is 150 mL/min.
Control the flow with the outlet valve on the flow meter (number 3 in
foam developing inside the flow chamber which could lead to erroneous oxygen measurement
values.
1. Valve to redirect sample for measurement 3. Flow meter with outlet valve
2. Sensor and flow chamber combination

3.9.3 Sensor insertion

Insert the sensor straight into the flow chamber or socket. Do not twist the sensor.
Hand tighten the attaching collar.
Connect the sensor cable.
Check for leaks; replace O-rings if product leaks are visible.
valves
pipe bends
the suction side of any pumps
a CO
injection system or similar
2
Figure 21 Typical measurement schematic
Figure 21
4. Drain
Installation
below) to avoid
33