Installation

3.2 Installation check list

For a complete installation, proceed to the following actions following the instructions in this
manual very carefully:
1. Install the 510 analyzer. Refer to
2. Install the sensor in the sample pipe or flow chamber and connect it to the instrument. Refer
to
3. Turn the instrument on and set the operating language. Refer to
4. Set the security levels, users ID's and passwords. Refer to
5. Verify that the sensor cap selected (i.e. low or high range: K1200-L or K1200-H)
corresponds to the sensor in use. Refer to
6. Enable thermal cutoff and set it to the minimum possible value (e.g. 10°C if the liquid in the
sample line is at 0-5°C). This is critical to ensure long-term stability. Refer to
page
7. Set the measurement interval according to your response time needs. Refer to
on page
8. Enable the calibration timer and set the number of days to 180 days (6 months) or
according to your quality needs to ensure a regular sensor calibration. Refer to
on page
9. Enable the service timer and set the number of days to 360 days (1 year) or according to
your quality needs to ensure a regular sensor service. Refer to
10. If required, program the relay and analog outputs to ensure an event is set when a
calibration or service is required. These can be programmed for each as custom events.
Refer to
instrument will automatically display this information as a warning message. Refer to the
table
As the system is factory calibrated and does not require an initial calibration. Refer to
Calibration Menu on page 53
for operation. If a problem occurs, refer to
overcome, please contact your Hach representative who will be happy to assist you.
20
section 3.9 on page
33.
50.
50.
91.
Section 8 on page
List of events and alarms on page
for more details. The instrument and sensor should now be ready
section 3.3 on page 21
section 6.2 on page
59. Note that when a calibration or service timer has expired, the
97.
Troubleshooting on page
through
section 3.8 on page
section 13.2 on page
Section 10 on page
83.
50.
section 6.2 on
section 6.2
section 13.2
section 13.2 on page
95. If the difficulty cannot be
32.
91.
91.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orbisphere 510

Table of Contents