Hach ORBISPHERE K1200 User Manual page 78

Analyzer
Hide thumbs Also See for ORBISPHERE K1200:
Table of Contents

Advertisement

Commands
The "Command Output Buffer" is formatted as follows:
Name
Output command toggle (OCT)
Output command ID (OCI)
Output command data byte 1 (OCD1)
Output command data byte 2 (OCD2)
Output command data byte 3 (OCD3)
Output command data byte 4 (OCD4)
The "Command Input Buffer" is located just after the measurement data and is formatted as
follows:
Name
Input command toggle (ICT)
Input command status (ICS)
Input command data byte 1 (ICD1)
Input command data byte 2 (ICD2)
Input command data byte 3 (ICD3)
Input command data byte 4 (ICD4)
The following commands are available:
Change product
Activate sensor (valid for EC sensors only)
Set channel hold status (valid for TC and LDO sensors only)
Change product command - output
Name
Value
OCT
1-2
OCI
1
OCD1
0
OCD2
0-99
OCD3
0-1
OCD4
Change product command - input
Name
Value
ICT
1-2
ICS
0-3
ICD1
ICD2
ICD3
ICD4
Output byte
Output byte
Output byte
Output byte
Output byte
Output byte
Input byte
Input byte
Input byte
Input byte
Input byte
Input byte
Command ID
Channel number:
0 = Channel 1
Product number
Erase measurement files:
0 = Never erase the measurement files.
1 = Erase measurement file if necessary (e.g. gas unit changes)
Not used
0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no or product number)
3 = Execution failure
Not used
Not used
Not used
Not used
Communication Menu
Type
Size
8 bits
8 bits
8 bits
8 bits
8 bits
8 bits
Type
Size
8 bits
8 bits
8 bits
8 bits
8 bits
8 bits
Comment
Comment
Offset
0
1
2
3
4
5
Offset
28
29
30
31
32
33
77

Hide quick links:

Advertisement

Table of Contents