Specifications
User interface
Sample temperature
Sample pressure
Sample flow rate requirement
Sample types
CIP
Measurement range
Repeatability
Reproducibility
Accuracy
Limit of detection (LOD)
Response time (90%)
Warm-up time
Display resolution
Calibration
Calibration sample
Accessories
In process sensor installation
devices
Ambient temperature
Humidity
Power supply
Wall and pipe mount
(H x D x W)
Panel mount: Face (housing)
(H x D x W)
K1200 28 mm sensor
Calibration device
14
Touch-screen panel
Password protection
Sensor
Measurement from -5 to 50°C (23 to 122°F)
Sensor resistant to temperature from -5 to 100°C (23 to 212°F)
1 to 20 bar absolute (14.5 to 290 psia)
50 to 300 mL/min
Water only
Not designed for use in CIP processes
Low range sensors: 0 to 2,000 ppb (dissolved. Indicatives values until 5000 ppb)
High range sensors: 0 to 40 ppm (dissolved)
Low range sensors: ± 0.4 ppb or 1%, whichever is the greater
High range sensors: ± 0.015 ppm or ± 2% whichever is the greater
Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater
Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater
Low range sensors: 0.6 ppb
High range sensors: 0.015 ppm
Low range sensors: < 10 seconds in gas phase; < 30 seconds in liquid phase
High range sensors: < 10 seconds in gas phase; < 50 seconds in liquid phase
None
0.1 ppb
Low range sensors: Single point calibration (zero)
High range sensors: Two at cap replacement (zero and air), one during use (air)
Low range sensors: Standard 99.999% N
High range sensors: Standard 99.999% N
Accessories
Active spots, spare sensors, tool kit, portable calibration setup, etc.
Ask your local Hach representative for more details on all available spare parts and
accessories.
Installation
ORBISPHERE insertion/extraction valve for installation on Varinline access units
ORBISPHERE 28 mm sensor weld-on socket
28 mm stationary housing for installation on Varinline access units
-5 to 50 °C (23 to 122 °F)
0 to 95% non-condensing relative humidity
Universal 85-264 VAC @ 50/60 Hz, 25 VA
10-36 VDC, 25 W
Weights
236.5 x 160 x 250 mm - weight 4.25 kg
9.31 x 6.30 x 9.84 in. - weight 8.82 lbs
156 (123) x 250 x 220 (214) mm - weight 3.35 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) in. - weight 6.62 lbs
0.74 kg (1.63 lbs)
0.7 kg (1.54 lbs)
Displays concentration, trend graph, diagnostics,
alarm status, historical data.
Five levels of authorised access to configuration and
data management
(quality 50) or equivalent oxygen free gas
2
(quality 30) or equivalent oxygen free gas, air
2