G
S
ENERAL
PECIFICATIONS
– A
I
SOLATION
NALOG INPUTS
Basic insulation, IEC 61010-1 (3
for secondary circuits derived from the Mains
Input channel to Ground
Input channel to channel
Impedance across barrier
C
M
R
R
OMMON
ODE
EJECTION
C
C
C
HANNEL TO
HANNEL
ROSSTALK
T
H
D
OTAL
ARMONIC
ISTORTION
Voltage Mode
±10.0 V
±1.0 V
±0.1 V
M
S
R
EASUREMENT
AMPLING
ATE
P
D
F
ROGRAMMABLE
IGITAL
ILTERS
None (No Filter)
High Performance (FIR)
Low Latency (CIC)
Medium Latency (CIC+CFIR)
Post Filter Blind Divider
C
C
P
HANNEL TO
HANNEL
HASE
P
R
ASSBAND
IPPLE
CLOCK
TRIGGER S
AND
E
S
LECTRICAL
PECIFICATIONS
Maximum Input Voltage
Input Impedance
Minimum Input Pulse Width
Maximum Output Frequency (CLK)
V
IL
V
IH
V
OL
V
OH
I
MAX
EX1401 Introduction
rd
): Pollution degree II, Material IIIa, Altitude < 5000m, Overvoltage Category II, applicable
±500 V Peak continuous working voltage
±1000 V Peak continuous working voltage between channels
100pF || Gas Discharge Tube rated for 600V
(CMRR), T
ATIO
YPICAL
10mV range
100mV range
1V range
10V range
The 10V input range is reduced by increasing common mode AC voltage above
100Vpk/60Hz.
, T
YPICAL
-145 dB < 1kHz
(THD), 100 H
Z
Typical: -90dB
Typical: -87dB
Typical: -85dB
Minimum: 0.000002328 Samples per Second, per channel
Maximum: 20,000 Samples per Second, per channel
Raw data
1 to 16 number of /2 stages (Selectable & Customizable)
/4 to /8192 (Selectable)
Low latency CIC filter, followed by /4 FIR Filter (Customizable)
1 - 65536 (selectable)
M
ATCHING
<±1.5 μS
<±0.005 dB over the passband of 0.45 * Sample Rate
PECIFICATIONS
-0.5V to 5.5V, ESD protected
50 Ohm series terminated with a 10k Ohm pulldown resistor
1 μs
10 MHz
< 0.8V
> 3.5V
< 0.65V @ 32 mA
> 3.0 V @ 32 mA
50mA max
DC
50/60 Hz
-140 dB
-140 dB
-140 dB
-140 dB
-140 dB
-140 dB
-140 dB
-140 dB
www.vtiinstruments.com
400 Hz
1000 Hz
-130 dB
-110 dB
-130 dB
-90 dB
-130 dB
-80 dB
-130 dB
-80 dB
19