ILX Lightwave LDC-3926339 User Manual page 98

6a current source module
Table of Contents

Advertisement

Table 5.7 Error Codes
Error Code
E-214
E-222
E-223
E-226
E-501
E-503
E-504
E-505
E-507
E-508
E-509
E-510
E-511
E-529
E-535
E-601
E-602
E-620
E-621
E-622
E-710
E-711
E-713
E-802
02_03
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Explanation
Response is too long to output.
Set value is over range.
Set value is under range.
Error in arbitrary block specification.
Interlock Open forces Laser output off.
Laser Voltage limit forces output off. "Open Circuit" error.
Laser Current Limit forces output off.
Laser Voltage Limit forces output off.
Laser Monitor Diode Power Limit forces output off.
External TEC Output Off Status forces LAS output off.
External TEC Temperature Limit forces LAS output off.
Out of Tolerance status forced LAS output off.
Laser Hardware Error forces output off. (e.g. low power line voltage).
Laser Output is off, but Laser status thought it was on.
Laser Mode changed while output on.
Internal error: recalled bin has incorrect checksum. (Settings do not match bin).
Internal error: task synchronization error.
Internal error: resource unavailable.
Internal error: message undeliverable to task.
Internal error: could not send message to mainframe.
AC Power Low Error detected
AC Power Low Error detected
Power Brown-Out Error detected
Calibration error: measurement entered before calibration was ready.
Troubleshooting Guide
LDC-3926339
5
C H A P T E R
81

Advertisement

Table of Contents
loading

Table of Contents