Laser Forward Voltage Measurement Calibration - ILX Lightwave LDC-3926339 User Manual

6a current source module
Table of Contents

Advertisement

C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
5
C H A P T E R
Local Calibration

Laser Forward Voltage Measurement Calibration

This procedure is for calibrating the laser forward voltage (LDV) measurement
circuits. The internal calibration procedure sets the current to two values; you
must enter the resulting forward voltage at both of these points.
1
2
3
3a
3b
3c
4
5
6
72
LDC-3926339
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
With the output off, connect a 1 Ω load resistor to the LASER output terminals (Laser
Anode, pin 9, and Laser Cathode, pin 5).
Connect a voltmeter with at least 15V range and accuracy of ±0.01 V across the load
resistor.
Press MAIN to enter the LDV calibration mode.
Press the Sys. Config soft key, then the Cal. soft key.
Press Cal LAS.
Press the LDI Cal softkey. The LDV calibration screen appears.
Press the Start softkey. The internal calibration procedure turns on the output and drives
the LD current to approximately 25% of full range. Allow the current to settle for at least 10
seconds.
Measure the voltage across the load resistor and enter the value in Volts (V) into the
VReal field and press ENTER.
Do not enter more than seven characters, including the decimal point. The internal
calibration procedure then drives the LD current to approximately 75% of full range. Allow
the current to settle for at least 10 seconds.
Measure the actual voltage across the load resistor, enter that value (in Volts) into the
highlighted "VReal" adjust field on the screen, and press ENTER.
After you have entered the second value, the internal calibration procedure calibrates the
laser forward voltage measurement circuit. The screen then displays a message
indicating that the calibration has finished.