C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
5
C H A P T E R
Remote Calibration
6
7
76
LDC-3926339
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Send the query LASER:CAL:STATUS? to make sure the internal calibration procedure is
ready for a measurement.
The calibration procedure is ready for a measurement if it returns the value 1. Query the
status until the value that is returned is 1.
Enter the measured voltage in Volts (V) by sending the command LASER:CAL:MEAS
<voltage in V>.
For example: LASER1:CAL:MEAS 4.429
After you have entered the second voltage measurement, the internal calibration
procedure performs the calibration. The calibration procedure sets the status to 2,
indicating the calibration is finished.