Adjusting The Sealing Parameters - Agilent Technologies PlateLoc User Manual

Thermal microplate sealer
Hide thumbs Also See for PlateLoc:
Table of Contents

Advertisement

80
Chapter 5: Optimizing seal quality
PlateLoc User Guide

Adjusting the sealing parameters

About this topic
Checking the seal
quality
This topic explains how to check the seal quality to determine how to
adjust the sealing parameters.
!! DAMAGE HAZARD !!
after the seal cycle is finished. The sealed plate might be hot.
During the optimization test, you should check the seal quality after
each seal cycle to determine the adjustments to make. For detailed
adjustment guidelines, see the PlateLoc Thermal Plate Sealer Seal
Selection Guide. You can locate the guide on the Velocity11 website at
www.velocity11.com.
To check the seal quality:
1. After the seal cycle is finished, wait for the plate to cool down
enough so that it is safe for handling.
Note: Some seals cannot be peeled easily if completely cooled.
2. Remove the plate from the PlateLoc.
3. Carefully peel off the seal material by lifting one of the corners of the
seal.
Peel the seal from either corner.
4. Inspect the underside of the removed seal material.
Good seal. If there are unbroken impressions of each well on the
underside of the seal material, then the plate was properly
sealed.
Do not handle the plate immediately

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents