Plateloc Diagnostics Parameters And Commands - Agilent Technologies PlateLoc User Manual

Thermal microplate sealer
Hide thumbs Also See for PlateLoc:
Table of Contents

Advertisement

120
Appendix B: Quick reference
PlateLoc User Guide

PlateLoc Diagnostics parameters and commands

Controls tab
Time Settings area
Parameter
Sealing time (sec)
Temperature Settings area
Parameter
Sealing temperature (°C)
Actual temperature (°C)
Status area
Status
Ready
Cooling down
Warming up
Cycle in progress
Cycle pending
In menu
Error
Description
The amount of time, in seconds, the hot plate
is in contact with the sealing material and the
plate.
Description
The temperature, in Celsius, of the hot plate
when the seal cycle starts.
The current temperature of the hot plate.
Description
The indicator that the PlateLoc is within 2 °C
of the specified sealing temperature and is
ready for a run.
The indicator that the PlateLoc is in the
process of cooling down to the specified
sealing temperature.
The indicator that the PlateLoc is in the
process of warming up to the specified
sealing temperature.
The indicator that the PlateLoc is in the
process of sealing a plate.
The indicator that shows the PlateLoc hot
plate is warming up or cooling down.
The indicator that another operator is setting
parameters or using controls at the touch
screen.
The indicator that one or more errors have
occurred.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents