The E5386A Half-channel Adapter (for use with the 16760A logic analyzer)

18
The E5386A Half-channel Adapter is intended to be used with the Keysight
16760A logic analyzer in half-channel state mode and supports the
E5402A, E5405B, and E5406A probes.
The E5386A Half-channel Adapter has its own ID code. When using the
adapter, the 16760A logic analyzer recognizes its code rather than that of
the probe which is attached to the target. Therefore, the user interface
format menu doesn't automatically set thresholds to the proper values.
You need to go into the threshold menu and select (differential, custom, or
standard settings).
When using the adapter in half-channel state mode:
Clock-bits are not available in half-channel state mode (although JCLK
on the master is still used).
Be sure to connect Master pod 1 of the logic analyzer to the upper bits,
8-15 + clk, on the half-channel adapter. This is necessary to connect
the clock in the system under test to the logic analyzer system clock.
Using the E5386A does not reduce the performance of the 16760A
logic analyzer and the soft touch probes.
If the E5386A is used in full-channel state mode, the thresholds on the
unused (odd) bits are floating. This could result in spurious activity
indicators in the format menu.
Keysight Technologies E5400-Pro Series Soft Touch Connectorless Probes User's Guide