00809-0100-4148
Table 4-1: Rosemount 148 Troubleshooting Chart
Symptom
Potential source
Loop wiring
Operation and Maintenance
(continued)
Corrective action
Check for adequate voltage to the
transmitter. It should be 12.0–42.4 V
dc (over entire 3.75–23 mA operating
range).
Check for shorts and multiple grounds.
Check for proper polarity at the signal
terminal.
Check the loop impedance.
Check wire insulation to detect
possible shorts to ground.
June 2019