Adash 4500 – VA5Pro
When graph contains data the averages and speed notification appear in the graph's top right corner. See the
graph 4. The 3/4 means the 3 of 4 averages has been taken and 17.7Hz is the speed on the beginning of the
reading. The light orange lines in the time signal are tacho marks.

Input Buffering

Measured signal is collected into the input memory. The analyze procedures read the signal data from the
memory and compute all required results, which are displayed on the screen. In the case, when too many
results are required, the computation is slow and displayed graphs are not in time - they are delayed. We are
talking about Real Time (RT) graphs (you see the actual results) or about Non Real Time (non RT) graphs
(delayed results).
This information is displayed in the right upper corner – Measuring; RT or Measuring; non RT.
After the measurement is stopped, the computation continues until the input memory is empty. The bellow
window appears.
You can wait for all data acquisition or if you are not interested of measurement from input memory press the
Stop button to stop data processing.
Graph Max/Min
It maximizes the selected graph to the whole screen or returns back to the initial screen with more graphs.
Graphs Control
Use up and down arrow button to change the selected graph. Use the left and right button to change the cursor
position. You can do it also by tapping the touch screen. If you tap the non selected graph, it becomes selected.