Probe Attenuation Ratio; Functional Test - Keysight Technologies N2750A User Manual

Differential probes
Table of Contents

Advertisement

1
Using the Probes
Refer to the oscilloscope's online help for information on Multipurpose functions.
No Action
Disables the probe button.

Probe Attenuation Ratio

The probe's attenuation ratio can be set to 2:1 or 10:1. When you attach the probe
to the Infiniium oscilloscope, the AutoProbe interface automatically sets the probe
attenuation to the value necessary to make the dynamic range of the probe
greater or equal to the level required to measure the current input signal.

Functional Test

Use the following generic measurement procedure to ensure that your probe is
functioning properly.
1 Connect the probe to an oscilloscope channel input and ensure the input
2 On the oscilloscope, click Utilities > Calibration Output to assign probe
3 Connect the probe tip to the oscilloscope's front-panel AUX Output connector.
4 Press the oscilloscope's [Auto Scale] button.
5 If the probe is functioning properly, a square wave should now be displayed on
22
impedance of the oscilloscope matches the output impedance of the probe
(50W).
compensation (Probe Comp) to the AUX Output connector.
the oscilloscope.
N2750/1/2A User's Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N2751aN2752a

Table of Contents