Performance Verification
Table 6
Required Test Equipment (Sheet 1 of 2)
Test Equipment
Infiniium Oscilloscope
Digital Multimeter
Vector Network Analyzer
(VNA)
Test Port
Cables/Adaptors
PV Fixture
Calibration Module
Probe Power Supply
Probe Adapter
60
Recommended Model
If a 90000 Q/X-Series oscilloscope is used, an N5442A precision BNC
adapter is required.
Keysight 33401A or equivalent). Critical specification: 2 wire resistance
accuracy
Keysight E8361A/C (2-port). Although
this procedure assumes an E8361A/C, an
equivalent VNA that covers at least a
50 MHz to 13 GHz range is acceptable.
The VNA needs to have a bias port for
PORT 1 which consists of internal bias
T's and a BNC port that allows bias to be
applied to PORT 1.
Two cables with male 3.5 mm connectors. If 2.4 mm or 1.85 mm cables are
selected, use Keysight 11901D adaptors to convert to 3.5 mm male
connectors.
E2655B or N5443A
Performance Verification
(PV) Fixture. This fixture is
included with the Deskew
Kit which is shipped with
Infiniium oscilloscopes.
Keysight N4691B Electronic Calibration
Module, 300 kHz to 26.5 GHz, 3.5 mm, 2-port.
An equivalent 3.5 mm cal kit can be
substituted for the N4691B.
Keysight 1143A Probe Offset Control and
Power Module
Keysight N1022B Probe Adapter. Or, use a Keysight N5477A
Sampling Scope Adapter with a Keysight N5442A Precision
BNC adapter.
Performance Test
1
2
?
?
?
?
?
?
?
?
N2750/1/2A User's Guide