Performance Verification
Figure 49
22 On the VNA, press [Trigger] > Single to trigger a single sweep.
23 Press [Scale] > Reference Level and adjust the reference level until the 50 MHz
24 The display should look similar to
70
Measurement Setup for V
point (at the display's left side) is at center screen.
NOTE
The reference level should be approximately –15.3 dB, but can vary a few
tenths of a dB either way.
NOTE
If it looks noticeably different, the probe tip wires may not be making contact
under the pinchers.
out
Figure 50
on page 71.
N2750/1/2A User's Guide