Performance Test Record - Keysight Technologies N2750A User Manual

Differential probes
Table of Contents

Advertisement

Performance Test Record

Table 7
N2750/1/2A Performance Test Record
Model #:
Serial #:
Recommended Test Interval: 1 year / 2000 hours
Probe Amplifier
Test 1. DC Input Resistance Performance Test
N2750/1/2A
Test 2. Bandwidth
N2750A
N2751A
N2752A
N2750/1/2A User's Guide
Date:
Recommended next test date:
Test Limits
200 kΩ ± 2% (differential,
between A and B leads)
100 kΩ ± 2% (single-ended,
between A and ground)
100 kΩ ± 2% (single-ended,
between B and ground)
≥ 1.5 GHz
≥ 3.5 GHz
≥ 6.0 GHz
Tested by:
Result
Performance Verification
Pass/Fail
6
73

Advertisement

Table of Contents
loading

This manual is also suitable for:

N2751aN2752a