APPENDIX A – DISSOLVED OXYGEN SENSOR
A.7.3 Membrane replacement (oxygen membrane cap)
Conditions which could require the membrane and electrolyte replacement are the following:
Calibration is not systematically achieved (try at least three times) and "OXYGEN SENSOR
CALIBRATION ERROR" appears on screen.
The oxygen sensor responds more slowly than usual or drifts.
The membrane of the cap is mechanically damaged and shows leakage, holes or scratches.
Read-out of over 0.2 ppm is displayed when carrying out the sensor check in the absence of
oxygen.
The oxygen sensor, filled with electrolyte, has been stored for a long time at temperatures outside
those recommended (- 10 to 40
A.7.4 Replacement of membrane can be done using the OXYGEN SENSOR MAINTENANCE KIT:
Locate and pull out the following parts from the maintenance kit:
one oxygen measuring "green" membrane;
one o-ring;
o-ring mounting tool.
Remove the protection ring from the membrane cap. Remove and discard the black o-ring and the
membrane(s).
Fit the new o-ring over the mounting tool and roll it down to the widest part of the tool.
Place the cap on a desk top with its narrow end facing up.
Position the measuring membrane "green" on top of the cap.
Place the widest part of the tool against the membrane. Slightly pressing the tool, slide the o-ring
carefully into the slot of the cap thus holding the membrane in position.
Cut away excessive membrane, with fine scissors, far from the o-ring to avoid damaging the
membrane.
Finally, recap the protective plastic ring.
A.7.5 Oxygen sensor cleaning
During the calibration procedure, the oxygen sensor current is shown:
Checking oxygen sensor:
Current: XXX nA. % of last calibration. XXX.X%
The acceptable current is 30-90 nA.
IDRONAUT – Brugherio (MB)
C).
o
OCEAN SEVEN 304Plus CTD
57
12-2015