idronaut Ocean Seven 304Plus Operator's Manual page 38

Ctd probe
Table of Contents

Advertisement

SECTION THREE – COMMANDS
Change the speed to the 9.6Kbps !
.......
Parameter Setup Id Name Code Mux Digits
01 Press 0000 254 0002
02 Temp 0001 254 0003
03 Cond 0002 253 0003
04 Sal 0004 255 0003
05 SoundV 0016 255 0004
CMD>I)nitialize,M)odify,D)elete,Q)uit
cmd>
When this prompt appears, type Q to exit and then switch off the OS304Plus CTD and proceed with
the next step. The CTD will wake using the new 9600 bps speed !
Set up the ITERM to the 9600 bps speed
Now, ITERM must be modified according to the firmware upgrading speed and set-up.
Note
 The ITERM port configuration is the following:
 The ITERM programme help contains the description of the programme main functions.
 At higher baud-rates, even if possible, it cannot be used to upgrade the OS304Plus CTD firmware.
Therefore, modify the OS304Plus communication speed and set it to the 9600bps value before
continuing/proceeding with the firmware upgrading.
 During the firmware upgrading, the Xon/Xoff RS232C protocol is used by the CTD to communicate with the
ITERM programme. Therefore, before proceeding, the Xon/Xoff must be enabled in the ITERM
programme by accessing the serial port set-up and enabling the "Software transmit" and "Software
receive".
IDRONAUT – Brugherio (MB)
Communication speed
Data bit
Stop bit
Parity
Flow Control
OCEAN SEVEN 304Plus CTD
9600 bps
8
1
None
Xon/Xoff
30
12-2015

Hide quick links:

Advertisement

Table of Contents
loading