idronaut Ocean Seven 304Plus Operator's Manual page 35

Ctd probe
Table of Contents

Advertisement

SECTION THREE – COMMANDS
02 Temp 0001 254 0003
03 Cond 0002 253 0003
04 Sal 0004 255 0003
05 UNKNW 0255 255 0000
06 UNKNW 0255 255 0000
07 UNKNW 0255 255 0002
08 UNKNW 0255 255 0002
09 UNKNW 0255 255 0002
10 UNKNW 0255 255 0002
11 UNKNW 0255 255 0002
12 UNKNW 0255 255 0002
13 UNKNW 0255 255 0002
14 UNKNW 0255 255 0002
15 UNKNW 0255 255 0002
CMD:I)nitialize,M)odify,D)elete,Q)uit
At the end of the list, commands available to perform the configuration are shown:
Initialize
Modify
Delete
Quit
The following parameters can be configured by means of the Modify command:
Sensor code[0..24,255=NU]
Sensor mux.[0..3,255=NU]
Sensor precision [0..6]
The below table reports the value associated to the available sensors/parameters:
Sensor Code
0
1
2
3
4
5
6
16
18
22
24
29
Mux channel 255
Mux channel 254/3
The present firmware release includes the following processed/derived parameters: SALINITY,
SOUND SPEED, DENSITY, CONDUCTIVITY@ 20°C, CONDUCTIVITY@ 25°C and TOTAL
IDRONAUT – Brugherio (MB)
Completely deletes the list.
Allows the operator to enter a new sensor/parameter in the list or modify an
existing one. See below for the details about configuration.
Allows the operator to delete a configured sensor/parameter from the list.
Terminates editing and ends the set-up too.
Sensor Description
Pressure
Temperature
Conductivity
Conductivity @20°C
Salinity
O
ppm
2
O
%
2
Sound-V
Conductivity @25°C
Density
Turbidity
TDS
= Derived parameter does not belong to a single sensor.
= Special CTD preamplifier.
OCEAN SEVEN 304Plus CTD
Sensor Mux.
Sensor Precision
254
254
253
255
255
255
1
255
255
255
0
255
2
3
3
3
3
2
1
4
3
6
1
3
12-2015