SECTION THREE – COMMANDS

3.2.6. Burst Sampling

Burst sampling allows collecting a "burst" of data at regular time intervals. During each single burst,
data is acquired using a configurable sampling rate. To save the internal battery energy, the CTD
switches OFF and ON by itself between bursts. Burst sampling is somehow complementary to "time"
acquisitions.
An example of a burst sampling cycle configuration follows:
OCEAN SEVEN 304Plus - ID:304-0010101 {USR}[9.0_00-02/2012] 16:43:44.77 11-07-2012
Burst measurements
Whole number of burst:5 <
Data set per single burst:1 <
Time between consecutive burts: 00:00:30[hh:mm:ss]<
Time between dataset in single burst [250ms..60000ms]:250 <
First acquisition time: 00:00:00[hh:mm:ss]<
Do you confirm the above setup [1(Yes,0(No] ?
Confirm ?:0 < 1
OpenLOG(5)..oK
BurstAcquisition..HF
Acquisition: <^C>Stop
Press Temp Cond Sal
0.34 26.5440 1.3686 1.2340 16:44:06.37M
Burst cycles left: 4
NextACQTime 16:44:35.39 11-07-2012
OS304Plus Shutdown
The burst data acquisition configuration parameters are described in section 4.
OCEAN SEVEN 304Plus - ID:304-0010101 {USR}[9.0_00-02/2012] 16:45:35.01 11-07-2012
Work.Mem:Cnf[oK].Calb[oK].Stat[oK].WakeUp[9F]
Data memory...[1 GByte]
WarmUp-
Battery..3.60 VDC....CTD..Ok Analogue:.Ok
OperatingMode...Burst
BurstAcquisition..HF
Acquisition: <^C>Stop
Press Temp Cond Sal
0.33 26.5543 1.3686 1.2340 16:45:39.04M
Burst cycles left: 1
NextACQTime 16:46:05.06 11-07-2012
OS304Plus Shutdown
WARNING
The configuration of zero or negative parameters could cause the CTD to behave in an unpredictable way.
IDRONAUT – Brugherio (MB)
OCEAN SEVEN 304Plus CTD
21