SECTION THREE – COMMANDS
The continuous cycle configuration parameters are described in section 4.0.
OCEAN SEVEN 304Plus - ID:304-0010101 {USR}[9.0_00-02/2012] 16:15:59.01 11-07-2012
Work.Mem:Cnf[oK].Calb[oK].Stat[oK].WakeUp[9F]
Data memory...[1 GByte]
WarmUp-
Battery..3.59 VDC....CTD..Ok Analogue:.Ok
OperatingMode...Continuous
OpenLOG(4)..oK
Acquisition: <^C>Stop
Press Temp Cond Sal
0.24 26.2479 1.3653 1.2340 16:16:02.75M
0.24 26.2479 1.3653 1.2340 16:16:03.04M
0.24 26.2483 1.3653 1.2340 16:16:03.31M
WARNING
The configuration of zeros or negative parameters could cause the CTD to behave in an unpredictable way.

3.2.5. Conditional Sampling

Output from a chosen sensor is monitored at configured regular time intervals. When it reaches the
configured boundary "Trigger", a "continuous sampling cycle" starts as configured. The cycle continues until
data from the chosen sensor falls back under the trigger. This data acquisition cycle is interrupted upon
receipt of the "Stop Acquisition" (CTRL-C).
Conditional sampling cycles can be performed by acting on the CTD ON/OFF switch. An example of
CTD configuration message follows:
OCEAN SEVEN 304Plus - ID:304-0010101 {USR}[9.0_00-02/2012] 16:19:16.45 11-07-2012
Conditional Sampling
Monitoring timeout: 00:00:30[hh:mm:ss]<
Sensor to monitor: 0)Pressure, 1)Temperature, 2)Conductivity
Select sensor:0 < 1
Sensor trigger value:0.100000 < 30
Data acquisition scan rate [125ms]:250 <
Do you confirm the above setup [1(Yes,0(No] ?
Confirm ?:0 < 1
Probe Switch-Off by itself
The next probe Wake-UP will start the configured acquisitions
NextACQTime 16:20:01.86 11-07-2012
OS304Plus Shutdown
The conditional profile configuration parameters are described in section 4.
OCEAN SEVEN 304Plus - ID:304-0010101 {USR}[9.0_00-02/2012] 16:21:01.01 11-07-2012
Work.Mem:Cnf[oK].Calb[oK].Stat[oK].WakeUp[9F]
Data memory...[1 GByte]
WarmUp-
Battery..3.60 VDC....CTD..Ok Analogue:.Ok
OperatingMode...Conditional
NextACQTime 16:21:31.74 11-07-2012
OS304Plus Shutdown
WARNING
The configuration of zero or negative parameters could cause the CTD to behave in an unpredictable way.