Shock Resistance; Vibration Resistance; Mechanical Load; Electromagnetic Compatibility (Emc) - Endress+Hauser Proline Promag P 200 Technical Information

Electromagnetic flow measuring system
Table of Contents

Advertisement

Shock resistance

Vibration resistance

Mechanical load

Electromagnetic compatibility
(EMC)

Medium temperature range

Conductivity

Pressure-temperature ratings

22
Sensor
IP66/67, type 4X enclosure
As per IEC/EN 60068-2-31
Acceleration up to 2 g following IEC 60068-2-6
• Protect the transmitter housing against mechanical effects, such as shock or impact.
• Never use the transmitter housing as a ladder or climbing aid.
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
Details are provided in the Declaration of Conformity.

Process

• –20 to +150 °C (–4 to +302 °F) for PFA
• –40 to +130 °C (–40 to +266 °F) for PTFE
T
[°F] [°C]
A
140
60
40
100
20
0
0
-20
-40
-40
T
Ambient temperature
A
T
Medium temperature
F
Area shaded in gray: the ambient temperature range of –10 to –40 °C (–14 to –40 °F) applies to stainless flanges
m
only
Harsh environment and IP68 only to +130 °C (+266 °F)
n
³ 20 mS/cm for liquids in general
The following material load diagrams refer to the entire device and not just the process connection.
n
PFA
m
PTFE
-40
-20
0
20
40 60 80 100
-40
0
100
Proline Promag P 200
120 140 160
180
[°C]
T
F
200
300
360
[°F]
Endress+Hauser
A0017724