Operation; Pumping Flammable/Pyrophoric Materials - Edwards RV3 Instruction Manual

Atex rotary vane pumps
Hide thumbs Also See for RV3:
Table of Contents

Advertisement

6

OPERATION

Note:
We recommend that you obtain and read the Vacuum Pump and Vacuum System Safety manual (publication
number P300-20-000), available from Edwards or your supplier.
If you supply your own control system you must include emergency stop devices. These should be
fitted with restart lockouts such that a new start command may only take effect after the lockouts
have intentionally been reset.
You must obey the instructions within this manual. Failure to do so will compromise the ATEX
compliance and may cause harm, injury or death to persons.
6.1

Pumping flammable/pyrophoric materials

You must obey the instructions and take note of the precautions given below, to ensure that
pumped gases do not enter their flammable ranges.
When flammable materials are present within the equipment you must:
Not allow air to enter the equipment.
Ensure the system is leak tight.
Use an inert gas purge (for example, a nitrogen purge) to dilute any flammable gases or vapours entering the
pump inlet, and/or use an inert gas purge to reduce the concentration of flammable gases or vapours in the
pump and in the exhaust pipeline to less than one quarter of the gases' published lower explosive limits
(LEL).
Use an inert gas purge in to the pump gas ballast connection to prevent the condensation of flammable vapours within
the pump mechanism and exhaust pipeline.
© Edwards Limited 2008. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
WARNING
WARNING
WARNING
A652-18-880 Issue C
Page 17