Other Important Messages - i-SENS CareSens N Voice Owner's Booklet

Hide thumbs Also See for CareSens N Voice:
Table of Contents

Advertisement

Message
What is said
"Error 5.
This error
Strip
message may
insertion
appear when
error. Please
the wrong
insert a
blood glucose
CareSens N
test strip is
strip
used instead of
properly."
CareSens N
blood glucose
test strip.
"Error 6
There is a
Meter
problem with
failure."
the meter.
"Error 8
An electronic
Meter
error occurred
failure."
during the test.
46 www.i-sens.com
What it means
What to do
Repeat the test
with a CareSens N
test strip.
Do not use the
meter. Contact your
authorised i-SENS
sales representative.
Repeat the test with
a new test strip. If
the error message
persists, contact
your authorised
i-SENS sales
representative.

Other Important Messages

Message
What is said
What it means
"The room
The
temperature
temperature
is too low. "
during testing
was below
the operating
range.
"The room
The
temperature
temperature
is too high. "
during testing
was above
the operating
range.
Note:
If the error messages persist, contact your authorised
i-SENS sales representative.
What to do
Move to an area
where the
temperature is
within the
operating range
(5-50˚C) and repeat
the test after the
meter and test strips
have reached a
temperature within
the operating range.
www.i-sens.com 47