NO
Interferent
17
L-Dopa
18
Maltose
19
Mannitol
20
Methyldopa
21
Pralidoxime Iodide
22
Salicylate
23
Tolazamide
24
Tolbutamide
25
Triglycerides
26
Uric acid
27
Xylose
User Performance Evaluation
A study evaluating glucose values from fingertip capillary blood
samples obtained by 100 lay persons showed the following
results:
100% within ± 0.83 mmol/L of the medical laboratory values
at glucose concentrations below 5.55 mmol/L and 97.4%
within ±15% of the medical laboratory values at glucose
concentrations at or above 5.55 mmol/L.
52 www.i-sens.com
Difference Averages
Interval 1
Interval 2
2.8~5.5 mmol/L 13.9~19.4 mmol/L
0.04 mmol/L
0.5%
-0.4 mmol/L
-1.1%
0.1 mmol/L
-0.7%
-0.03 mmol/L
0.2%
0.0 mmol/L
1.4%
0.1 mmol/L
-0.1%
-0.3 mmol/L
-2.8%
-0.3 mmol/L
-7.3%
-0.1 mmol/L
4.7%
-0.1 mmol/L
0.5%
-0.04 mmol/L
-1.0%

Warranty Information

Manufacturer's Warranty
i-SENS, Inc. warrants that the CareSens N Voice Meter shall be
free of defects in material and workmanship in normal use for
a period of 5 years. The meter must have been subjected to
normal use. The warranty does not cover improper handling,
tampering, use, or service of the meter. Any claim must be
made within the warranty period.
i-SENS will, at its discretion, repair or replace a defective meter
or meter part that is covered by this warranty. As a matter
of warranty policy, i-SENS will not reimburse the consumer's
purchase price.
Obtaining Warranty Service
To obtain warranty service, you must return the defective
meter or meter part along with proof of purchase to your
nearest i-SENS sales or customer service representative.
For Australian Consumers
Our goods come with guarantees that cannot be excluded
under the Australian Consumer Law. You are entitled to a
replacement or refund for a major failure and compensation
for any other reasonable foreseeable loss or damage. You are
also entitled to have the goods repaired or replaced if the
goods fail to be of acceptable quality and the failure does not
amount to a major failure.
Australian Distributor:
Pharmaco (Australia) Ltd
Gordon, NSW 2072
Phone: 1800 114 610
Website: www.caresens.com.au
NZ Distributor:
Pharmaco (NZ) Ltd
Auckland
Phone: 0800 GLUCOSE (0800 458267)