3-9 SYSTEM & NETWORK I/O MODULE CON-
TROLS (SETUP) – SYSTEM SIO
This menu provides access to several sub-
menus for setting parameters of the SIO (Sig-
nal Input/Output) and DIO (Digital
Input/Output) of the analyzer.
Analyzer and I/O, expert controls &
System & network I/O module controls...
CLD
-- System & Network I/O Module Controls --
System SIO module...
System DIO module...
Measure
Figure 3-39. System & Network I/O Module
Controls Menu
Press the ↵ or → softkeys to change to the
desired submenu.
If there is no SIO module installed in the ana-
lyzer, a corresponding message will be dis-
played instead of the menu.
The "System SIO Module" menu provides
submenus for setting up the output configura-
tions of the SIO signals. The SIO board can
contain 2 to 8 analog outputs, a serial inter-
face (RS232 or RS485), and three relay out-
puts. General configuration of the SIO board
is contained in its own manual. If the SIO
board is installed in the analyzer, the line
"Module installed:" must be set to "Yes."
To access this menu, in "System & network
I/O module controls...", select "System SIO
module...".
3-26
Operation
Main Menu
setup...
7.50 ppm
<<<
Back...
>>>
CLD

Analog output setup...

Serial interface setup...
Relay outputs setup...
Module installed:
Measure

Figure 3-40. System SIO Module Menu

Select a line with the ↑ or ↓ softkeys.
Select the variable or change to the submenu
with the ↵ or → softkeys.
Select the variable parameter with the ↑ or ↓
softkeys.
Confirm the new value with the Enter ↵ soft-
key or cancel and return to the last value with
the F2 softkey.
a.
Analog Output Setup
In the System SIO Module menu (Figure
3-40, page 3-26), select Analog output
setup...
CLD
Measure

Figure 3-41. Analyzer Modules Menu

Output number:
Choose the desired analog output (1-8) to
set the parameters. The number of out-
puts depends on the analyzer configura-
tion as 2, 4, 6, or 8.
Rosemount Analytical Inc.
A Division of Emerson Process Management
Model NGA2000 CLD
-- System SIO Module --
Back...
-- Analyzer Modules --
<<<
Back...
7.50 ppm
Yes
7.50 ppm
CLD: 1.0
MLT/CH1
MLT/CH2
MLT/CH3
>>>