Summary of Contents for Emerson Heated Flame Ionization Detector Module Model NGA 2000 HFID
Page 1
Instruction Manual 760004-A February 2002 Model NGA 2000 HFID Heated Flame Ionization Detector Module http://www.processanalytic.com...
Page 2
Teflon is a Registered Trademark of E.I. duPont de Nemours and Co., Inc. Kynar is a Registered Trademark of Atochem North America, Inc. SNOOP is a registered trademark of NUPRO Co. Emerson Process Management Rosemount Analytical Inc. Process Analytic Division 1201 N.
Safety System........................3-5 MAINTENANCE AND SERVICE ..................4-1 Overview..........................4-1 Oven ............................4-3 a. Removal ..........................4-3 b. Disassembly........................4-4 Burner ............................4-6 a. Temperature Sensor .......................4-6 b. RTD Detector ........................4-6 c. Igniter ..........................4-6 d. Flameout Sensor......................4-6 Rosemount Analytical Inc. A Division of Emerson Process Management Contents...
Page 4
Noise ..........................5-2 REPLACEMENT PARTS ......................6-1 Matrix .............................6-1 General ..........................6-2 Pneumatics ..........................6-2 Oven Components.........................6-3 RETURN OF MATERIAL ......................7-1 Return Of Material .........................7-1 Customer Service ........................7-1 Training..........................7-1 APPENDIX A - MENU DISPLAYS..................8-1 Contents Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 5
Figure 4-20. Front Panel – Exploded View ..................4-23 Figure 4-21. Accessing Front Panel Components ................4-24 Figure 4-22. Rear Panel Components ....................4-27 LIST OF TABLES Table 3-1. HFID Analyzer Module Alarms ..................3-8 Rosemount Analytical Inc. A Division of Emerson Process Management Contents...
Page 6
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 Contents Rosemount Analytical Inc. A Division of Emerson Process Management...
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effective- ness. NOTE Highlights an essential operating procedure, condition or statement. Rosemount Analytical Inc. A Division of Emerson Process Management Preface...
The internal compartment of the oven is vented to the main enclosure by the top and bottom vents. DO NOT RESTRICT THOSE VENTS. Consult the factory if flammable samples will be measured. Preface Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 9
PURGE AIR OUT outlet of the Analyzer Module because this may affect the sealing performance of the module. CAUTION. PRESSURIZED GAS This module requires calibration with a known standard gas. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5. Rosemount Analytical Inc. A Division of Emerson Process Management Preface...
Page 10
Power shall not be restored after enclosure has been opened (or loss of purge) until enclosure has been purged for a minimum of 6 (six) minutes at the minimum pressure of 689 hPa (10 psig). Preface Rosemount Analytical Inc. A Division of Emerson Process Management...
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder. Rosemount Analytical Inc. A Division of Emerson Process Management Preface...
This product complies with the standard level of NAMUR EMC. Recommendation (May 1993). NAMUR This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand. N 9 6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management...
The measured species concentration value from an Analyzer Module. Secondary Variable Data placed on the network by a module regarding current status, e.g., sample flow, source voltage and other diagnostic information. Rosemount Analytical Inc. A Division of Emerson Process Management Preface...
Page 14
The five function softkeys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software. System Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s). Preface Rosemount Analytical Inc. A Division of Emerson Process Management...
This measure of concentration is placed on the acquisition device. Negative Ion Collection Plate +90V Signal Conditioning Positive Carbon Ions Sample Fuel Figure 1-1. Flame Ionization Detection Technology Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management...
Burner Air Solenoid Valve Purge Air Regulator Burner Air & Fuel Regulators (Stacked Vertically) Network & Power Module Figure 1-2. Heated Flame Ionization Detector Analyzer Module - Top View Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications...
Materials in Contact With Sample . Stainless steel and glass-filled Teflon Dimensions........See Outline and Mounting Dimensions, Figure 2-5 Weight ........... 15.9 kg (35 lbs.) Mounting........Horizontally, custom-installed in a panel Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management...
Burner Air ........Zero-grade air Flow rate ........ 350 to 400 mL/min. THC ........≤1 ppm, CH Supply pressure..... 1723 to 3446 hPa-gauge (25 to 50 psig) Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications...
See the Preface section of the Platform Components manual for specifications regarding Platform-related components and the Preface of the I/O Module manual for specifications regarding I/O (e.g., relay outputs). Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management...
95% relative humidity, but not in ex- PURGE IN/OUT and SAMPLE BYPASS) is cess of 45°C wet bulb temperature. strongly recommended, preferably pre- cleaned, stainless steel, gas chromato- Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications...
SAMPLE IN HEATED SAMPLE BYPASS OUT ATTENTION Danger d'electrocution, d'explosion ou d'emanation de gaz toxique. Se refere aux details inscrits surla face du module. Figure 2-1. Back Panel Connections Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management...
ANSI/NFPA 496 standards. Calibration method and gases depend on the operating range, and the desired measurement accuracy. In all methods, zero and span gases are used, and are Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications...
Burner Fuel Pressure should be: 3101 to a test procedure chosen by the user). 3450 hPa-gauge (45 to 50 psig) for cylinder regulator, 1723 hPa-gauge (25 psig) nomi- nal for internal pressure. Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management...
1/4T - 1/4MPT RUN TEE BRASS BRASS 1/4T - 1/4MPT PRESSURE BRASS REGULATOR, FIXED FLOW BURNER AIR CONTROLLER, 0 - 60 PSIG REGULATED AIR Figure 2-2. Flow Diagram Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications...
T 6A peut entrainer une explosion provoquant des blessure graves 250 V ou mortelles. IGNITE Figure 2-4. Front Panel Connections, Controls and Indicators Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management...
The pres- clean. sure should be constant. • The burner exhaust tube must be slanted down a minimum of 6 degrees from horizontal. Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications...
Page 30
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 2-10 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management...
• RAM test - Checks the RAM on the 2. Turn ON the purge gas only. Perform a Analysis Computer PCB. leak check. Wait a minimum of 6 min- utes. Rosemount Analytical Inc. A Division of Emerson Process Management Operation...
Page 32
69 hPa-gauge (1 psig), and recheck the Test Results menu. Purge Gas variable until it reads "ON." DO NOT EXCEED 1378 hPa-GAUGE (20 PSIG). If the maximum setting is reached, Operation Rosemount Analytical Inc. A Division of Emerson Process Management...
System Set Up portion of the Technical sary). Configuration Menu in software. • If the flame has been lit, but the flame temperature increases slowly, per- form the following steps: Rosemount Analytical Inc. A Division of Emerson Process Management Operation...
4. Verify the capillary type in the Analyzer To remedy the problem, do the following: Manufacturing Data menu (under the Technical Configuration menu structure, select the following from the Main Menu: Operation Rosemount Analytical Inc. A Division of Emerson Process Management...
Flow rate for purge air in is at least the particular hydrocarbon present in the span 16 L/min. and case pressure is greater gas. Note that readings obtained during op- Rosemount Analytical Inc. A Division of Emerson Process Management Operation...
Page 36
6 min. The elapsed time can be monitored in the Technical Startup Analyzer menu. (Moni- NOTE tor the Purge Air Green LED (ON), Purge Do not over-torque rear panel screws. Operation Rosemount Analytical Inc. A Division of Emerson Process Management...
SAMPLE: 100 ppm CH in N at 3.5 psig (241 hPa) 1032 1376 1726 2064 psig AIR PRESSURE Figure 3-2. Typical Curves of Module Response vs. Pressure Setting on Air Pressure Regulator Rosemount Analytical Inc. A Division of Emerson Process Management Operation...
Screw #6, 4 to 5 turns Repeat torque sequence until all screws are tight. The gasket must fill in between the front panel plate and the en- closure. Figure 3-3. Front Panel Torque Sequence Operation Rosemount Analytical Inc. A Division of Emerson Process Management...
A T T E N T I O N Cover/Rear Panel Secur- Securing Front Panel Securing ing Screws (6) Screw (4) Screws (6) Figure 4-1. Removal of Cover and Insulation Shield Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service...
DC Power Supply Mod- ule Assembly 657413 Oven Assembly 659551 Fan Assembly 657414 Fixed Flow Controller Assembly 657434 Front Panel Assembly 656949 Figure 4-2. Locations of Major Assemblies of the HFID Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
Lift oven assembly from ana- three gas lines and seven electrical ca- lyzer. bles, noting location of mating connectors for re-installation. Oven Assembly 659551 Figure 4-3. Removal of Oven from Chassis Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service...
8. Lift the burner/thermal block up and PREAMP CONNECTORS out, while disconnecting exhaust. The electrical preamp connectors are frag- ile, handle with care to avoid breaking sol- Reverse procedure for installation. der connection. Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
Bridge 904294 Heated Bridge 659555 Heater/Cable Assembly 659643 Jam Nuts 659550 Insulating Washers 073737 For clarity, outer oven not shown in Figures B and C. Figure 4-4. Oven Assembly Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service...
Do not over- nector, note location. tighten! 4. Gently grasp RTD detector wires and 5. Re-attach wiring connector. pull out of hole. 6. Install oven cover. Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 45
The components shown can be replaced without removing burner/thermal block from oven. Oven not shown for clarity. Thermal block shown in phantom for clarity. Figure 4-5. Burner - Sensor, Flameout Detector, RTD Detector and Igniter Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service...
Disassembly of Burner/Thermal Block mal block. 1. Remove oven from analyzer module per Section 4-2a. Burner Sample capillary input Thermal Block Sample capillary Figure 4-6. Burner/Thermal Block Disassembly Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
All tools used for main- tenance must be free of contaminates. Combustion Burner Cap Chamber Retainer Burner Cap O-Ring 904373 Gasket 656931 Burner Base Burner Base Figure 4-7. Burner Disassembly Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service...
Sample Jet 657005 Position with tab over the air hole (approximately 1/16” to the right). Figure 4-8. Burner Jets 4-10 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
7. See Figure 4-7A. Insert new gasket and tighten. on combustion chamber, install burner cap and burner cap retainer, 3. Tighten jet nut. torque screws to 6 inch lbs. Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-11...
3. Refer to Figure 4-9. Remove the two fit. retaining screws, pull thermostat out. 8. Install cover. 4. Install replacement thermostat, attach with the two retaining screws. Capillary Cover Figure 4-10. Thermal Block Assembly Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-13...
2. Lay electronics assembly on bench, Computer Analysis Board do not disconnect cables or tubing. Preamp Assembly Electronics Assembly 656943 Screw Hex Nut Figure 4-11. Removing Electronics Assembly from Chassis 4-14 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
One at a time, remove the wiring con- nectors and attach to replacement board. Per Section 4-6, remove securing hard- ware from electronics assembly and lay on bench. Mount replacement board to electronics assembly. Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-15...
6. Re-assemble to signal board mount- ing screw. 3. Remove screw securing cable clamp holder to signal board. Center sen- sor in cable clamp Figure 4-13. Case Sensor Installation 4-16 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
Case Pressure Purge Switch 903690 The bracket does not have to be removed from the electronics assembly for this procedure. Figure 4-14. Case Pressure Purge Switch Installation Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-17...
3. Remove the two screws and washers from the top bracket and slide the preamp assembly out. Top Bracket Preamp Assembly 656945 Bottom Bracket Figure 4-15. Preamp Assembly Installation 4-18 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
1. Disconnect and note location of cables. 2. Remove the two hex nuts securing the fan to the chassis, lift fan assembly out. Fan Assembly 657414 Figure 4-16. Fan Assembly Installation Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-19...
2. Remove the four hex nuts securing the flow controller assembly to the analyzer module chassis. Flow Controller Assembly 657434 Figure 4-17. Flow Controller Replacement 4-20 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
FLOW Remove red cap after installation Regulated Air Out Sample Sensor 656418 Remove and discard bracket supplied with regulator, assembly as shown. Figure 4-18. Flow Controller Assembly Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-21...
Disconnect and note location of all wiring to DC power supply module. Remove the two hex nuts securing module to chassis, remove module. DC Power Supply Module Figure 4-19. DC Power Supply Module Replacement 4-22 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
O-Ring 657053 656444 011167 Sensor Fittings 657412 Burner Air Regulator 902832 Fuel Regulator 902832 O-Rings 008025 Toggle Switch Seal 898980 Figure 4-20. Front Panel – Exploded View Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-23...
3. Remove the purge air regulator mounting bracket screws. Front panel mounting screw Front panel (opposite side also) mounting screws Figure 4-21. Accessing Front Panel Components 4-24 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
4. The replacement regulator comes panel from the inside. with two panel mounting nuts, remove both and discard one of them. 5. Install switch seal. 6. Re-connect wiring connector. Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-25...
4. Install replacement flow switch. placement restrictor, install into solenoid. 5. Install purge switch onto tee fitting. 3. Re-connect tubes to restrictor, elbow 6. Re-connect tubes. and tee fitting. 4-26 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
Filter, Burner Air In 017154 Restrictor, Heated Sample Bypass Out – 10 Micron 619615 Restrictor, Heated Sample In – 40 Micron 619616 Figure 4-22. Rear Panel Components Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-27...
Figure 4-22. 2. Insert a clean, rigid piece of tube or rod (smaller than .25 inch diameter) 4-28 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management...
16. Verify oven temperature is greater greater than 5.5 psig. than 85°C. 16. Verify the purge gas switch has been 17. Verify the reference thermistor is activated. 100K ohm ±15% at 25°C. Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting...
11. Verify the collector wires are routed away from the heater. 11. Verify burner has been on and stabi- lized. 12. Verify the collector wires are clean and not damaged. 12. Check for gas leaks. Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management...
None Standard (3 ft LON and Pwr AM to Platform) System (10 ft LON and Pwr AM to 30A PS) Code Special Requirements None Customer Option Special HFID Example Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts...
Burner Air 3-Way Solenoid Valve 656218 Fuel In Solenoid Valve 903690 Case Pressure Purge Switch 656533 Purge Air Flow Switch 903728 Regulated Air In Check Valve 903647 Case Pressure Relief Valve Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management...
Restrictor, Heated Bypass Sample Out – 10 Microns 659619 Heated Sample In Restrictor Assembly – 40 Microns 659616 Restrictor, Heated Sample In – 40 Microns 657065 Thermostat 450°F 659643 Cartridge Heater Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts...
Page 72
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management...
Customer Service Center 1-800-433-6076 If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the condi- Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material...
Page 74
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management...
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 SECTION 8 APPENDIX A - MENU DISPLAYS Menu: 0 ANALOP Menu: 1 ANALSET MENU: 2 FLOCHEK Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays...
Page 76
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 3 FLOCHEKI1 MENU: 4 ACALSET MENU: 5 APARLST Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 77
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 6 ANALSETI1 MENU: 7 CALLIST MENU: 8 CALLISTI1 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays...
Page 78
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 9 ACALSETI1 MENU: 10 APARLSTI1 MENU: 11 AMMAN Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 79
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 12 AMMANI1 MENU: 13 AMSVC MENU: 14 AMSVCI1 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays...
Page 80
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 15 ADIAG MENU: 16 AMPWR MENU: 17 AM1V Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 81
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 18 AMTEMP MENU: 19 AMMISC MENU: 20 AMTREND Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays...
Page 82
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 21 ADIAGI1 MENU: 22 RANGESETAM MENU: 23 RANGESSETI1 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 83
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 24 LINRANGE1 MENU: 25 LINRANGE2 MENU: 26 LINRANGE3 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays...
Page 84
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 27 LINRANGE4 MENU: 28 LINRANGE0 MENU: 29 AMPWRI1 8-10 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 85
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 30 FLOCHEK1I1 MENU: 31 FILTER MENU: 32 AM1VI1 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-11...
Page 86
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 33 AMTEMPI1 MENU: 34 AM2VA MENU: 35 PLIMITSA 8-12 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 87
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 36 TLIMITSA MENU: 37 AMMISCI1 MENU: 38 ANALSIMPLE Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-13...
Page 88
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 39 FILTERI1 MENU: 40 LINRANGE0I1 MENU: 41 PLIMITSAI1 8-14 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 89
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 42 CALFACTORS MENU: 43 R1FACTORS MENU: 44 RN2FACTORS Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-15...
Page 90
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 45 RN3FACTORS MENU: 46 RN4FACTORS MENU: 47 AMHELPINDEX 8-16 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 91
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 48 LINRANGE1I1 MENU: 49 CALFACTORSI1 MENU: 50 APARLST2 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-17...
Page 92
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 51 APARLST4 MENU: 52 APARLST5 MENU: 53 APARLST6 8-18 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 93
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 54 IGNITION MENU: 55 LISTNOTES MENU: 56 LIGHTFLAMEI1 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-19...
Page 94
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 57 AUTOFLAMEI1 MENU: 58 DISPLAY MENU: 59 MPARMS 8-20 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 95
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 60 AMTOPINFO MENU: 61 ANALSET2 MENU: 62 MPARMSI1 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-21...
Page 96
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 63 TLIMITSAI1 MENU: 64 IGNITIONI1 MENU: 65 SELFTEST 8-22 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 97
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 66 AMHELPINDEX2 MENU: 67 SOFT_DIAG MENU: 68 CALI1 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-23...
Page 98
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 69 AM2VC MENU: 70 AM2VD MENU: 71 OVENTEMP 8-24 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 99
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 72 LINFUNCT MENU: 73 POLYSETUP MENU: 74 MIDPOINT1 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-25...
Page 100
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 75 POLYGAS1 MENU: 76 POLYSETI1 MENU: 77 POLYGAS2 8-26 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 101
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 78 POLYGAS3 MENU: 79 POLYGAS4 MENU: 80 POLYGAS5 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-27...
Page 102
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 81 POLYGAS6 MENU: 82 POLYGAS7 MENU: 83 POLYGAS8 8-28 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 103
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 84 POLYGAS9 MENU: 85 POLYGAS0 MENU: 86 MIDPOINT2 Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-29...
Page 104
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 87 MIDPOINT3 MENU: 88 MIDPOINT4 MENU: 89 LIGHTFLAME 8-30 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 105
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 90 EXP_CAL MENU: 91 ZEROI2 MENU: 92 EXP_CAL_DAT Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-31...
Page 106
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 93 EXP_CAL_DATI MENU: 94 UNITS MENU: 95 UNITSI1 8-32 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 107
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 96 POLYSETI2 MENU: 97 POLYSETI3 MENU: 98 RESET Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-33...
Page 108
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 99 STORE MENU: 100 ANALOPI1A MENU: 101 RFHIST1A 8-34 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 109
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 102 RFACTORSIA MENU: 103 RFHIST2A MENU: 104 RFHIST3A Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-35...
Page 110
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 105 RFHIST4A MENU: 106 SW_DIAGI1 MENU: 107 TWEAKI1 8-36 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 111
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 108 ANALSETI3 MENU: 109 STOREDPVA MENU: 110 ZEROI2A Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-37...
Page 112
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 111 ZEROI1A MENU: 112 ZERO_NOW2 MENU: 113 SPAN_NOW2 8-38 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 113
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 114 AMMISC2 MENU: 115 MPARMS2 MENU: 116 AUTOFLAME Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-39...
Page 114
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 117 REBOOT MENU: 118 ABOUT MENU: 119 ABOUT1 8-40 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 115
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 MENU: 120 ALARM1 MENU: 121 MANDATA Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A – Menu Displays 8-41...
Page 116
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 8-42 Appendix A - Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management...
Page 117
WARRANTY Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller.
Page 118
Instruction Manual 760004-A Model NGA 2000 HFID February 2002 Emerson Process Management Rosemount Analytical Inc. Fisher-Rosemount GmbH & Co. Process Analytic Division Industriestrasse 1 63594 Hasselroth 1201 N. Main St. Germany Orrville, OH 44667-0901 T (330) 682-9010 T 49-6055-884 0...
Need help?
Do you have a question about the Heated Flame Ionization Detector Module Model NGA 2000 HFID and is the answer not in the manual?
Questions and answers