Response Time; Automatic Range Change; Figure 3-24. Automatic Range Control Menu - Emerson Rosemount Analytical NGA2000 CLD Instruction Manual

Chemiliuminescence analyzer module
Table of Contents

Advertisement

Model NGA2000 CLD
Make sure that the corrections are not
excessive. If the correction is too ex-
cessive, the calibration routine will fail
and calibration of the analyzer will not
be possible.
f.

Response Time

The response time menus allows the set-
ting of the primary analyzer variable t90
times, the LON update rate and the output
delay time.
Analyzer and I/O expert controls &
Analyzer module setup...
Gas measurement parameters...
Response time/delay parameters...
CLD
Response time/delay Parameters
Range 1 t90 time:
Range 2 t90 time:
Range 3 t90 time:
Range 4 t90 time:
LON update time:
Output delay time:
ESCAPE
HOME
Figure 3-23. Response time/delay
The t90 time adjusts the filtering or damp-
ing factor for the concentration outputs for
each range. They are adjustable from 0.1
to 30 seconds with a default of 3.0 sec-
onds.
The LON Update Rate is the rate at
which the analyzer communicates over
the system network. It is adjustable to
"ASAP" (the fastest rate that the network
communicates), "10 per sec," and "1 per
sec."
Rosemount Analytical Inc.
A Division of Emerson Process Management
CAUTION.
Main Menu
setup...
7.50 ppm
10 per sec
Parameters
The Output Delay Time establishes the
delay for the DIO and analog outputs to
respond to a change in concentration
value. It is adjustable from 0.0 to 30.0
seconds.
To change a value, move the cursor to
the desired line and press the ↵ softkey.
Change the value using the ↑ and ↓ soft-
keys and then press the ↵ softkey again
to save the selection. Press the Back (F2)
softkey before pressing the ↵ softkey to
restore the previous value.
g.

Automatic Range Change

This menu function allows the setting and
enabling of the automatic range change
for each of the four ranges. The automatic
range switching for increasing values is
effective at the upper range limit of each
range, while for decreasing values it is the
next lower range limit times the hysteresis
percent times the upper range limit.
3.0 s
3.0 s
3.0 s
3.0 s
0.0 s
Automatic range change parameters...
CLD
Actual switch levels...
Switch level hysteresis:
Usage of range –1:
Usage of range –2:
Usage of range –3:
Usage of range –4:
Automatic range change control:
Absolute range upper limit:
Absolute range lower limit:
Measure
Figure 3-24. Automatic Range Control
Set the desired hysteresis level in the
range of 10% to 50% which is applied to
each range transition. (20% is the default
setting.) Enable the desired ranges and
enable the automatic range change con-
trol. To view the actual switching levels,
Instruction Manual
November 2002
Main Menu
Analyzer and I/O expert controls &
setup...
Analyzer module setup...
Gas measurement parameters...
-- Automatic Range Control --
Menu
Operation
760000-C
7.50 ppm
20 %
Enabled
Enabled
Enabled
Enabled
Enabled
10000 ppm
10.0 ppm
Back...
3-17