Emerson Rosemount Analytical NGA2000 CLD Instruction Manual page 11

Chemiliuminescence analyzer module
Table of Contents

Advertisement

Model NGA2000 CLD
Tampering or unauthorized substitution of components may adversely affect safety of this product.
Use only factory documented components for repair.
This analyzer module may tip instrument over if it is pulled out too far and the Platform is not prop-
erly supported.
Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at
the ultraviolet source. Use of ultraviolet filtering glasses is recommended.
The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure.
Mercury is highly toxic if absorbed through the skin or ingested, or if vapors are inhaled.
Handle lamp assembly with extreme care. If the lamp is broken, avoid skin contact and inhalation
in the area of the lamp or the mercury spill.
Immediately clean up and dispose of the mercury spill and lamp residue as follows:
Wearing rubber gloves and goggles, collect all droplets of mercury by means of a suction pump
and aspirator bottle with a long capillary tube. (Alternately, a commercially available mercury
spill clean-up kit, such as J.T. Baker product No. 4439-01, is recommended.)
Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all
mercury, lamp residue and debris into a plastic bottle which cab be tightly capped.
Label and return to hazardous material reclamation center. Do not place in the trash, incinerate
or flush down the sewer.
Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and sul-
fur dust
Rosemount Analytical Inc.
PARTS INTEGRITY
OVERBALANCE HAZARD
INTERNAL ULTRAVIOLET LIGHT
TOXIC CHEMICAL HAZARD
A Division of Emerson Process Management
WARNING.
WARNING.
WARNING.
WARNING.
Instruction Manual
760000-C
November 2002
Preface
P-3