5-1 LEAKS

Liberally cover all fittings, seals, and other
possible sources of leakage with a suitable
leak test liquid such as SNOOP (part
837801). Bubbling or foaming indicates leak-
age. Checking for bubbles will locate most
leaks but could miss some, as some areas
are inaccessible to the application of SNOOP.
For positive assurance that system is leak
free, perform one of the tests above.