Figure 3-56. Power Supply Voltages Menu - Emerson Rosemount Analytical NGA2000 CLD Instruction Manual

Chemiliuminescence analyzer module
Table of Contents

Advertisement

Model NGA2000 CLD
WARNING
FAILURE
SAFETY FAILURE
ANY
Power supply voltages...
CLD
-- Analyzer Diagnostics --
Power supply voltages
+15V analog is:
+15V analog was:
-15V analog is:
-15V analog was:
+5V digital is:
+5V digital was:
+25V power is:
+25V power was:
+12V analog is:
+12V analog was:
ESCAPE
HOME

Figure 3-56. Power Supply Voltages Menu

The "is" lines show the current value of
the power supplies. The "was" values
show the values when the analyzer was
manufactured. Changes of more than a
few percent should be tracked. The 24V
power supply may differ substantially if
the Rosemount Analytical power supply is
not used. Use the MORE (3) softkey to
see additional values.
Primary variable parameters...
CLD
Primary variables parameters
Raw measurement signal:
Signal gain setting:
Current range:
Barometric pressure compensation:
Pk-pk noise:
Peltier device voltage:
Reference duty cycle:
Duty cycle coefficient:
(Temperature compensation coefficients...)
Calibration factors...
ESCAPE
HOME
Figure 3-57. Primary Variable Parameters
Shows the value of internal parameters
used in the primary variable calculation.
Barometric pressure compensation may
be enabled if another analyzer has a
pressure measurement device and is set
to report its reading.
Rosemount Analytical Inc.
A Division of Emerson Process Management
7.50 ppm
15.06 V
0.0 V
15.26 V
0.0 V
4.980 V
0.0 V
23.73 V
0.0 V
11.98 V
0.0 V
MORE
INFO
7.50 ppm
521651.7
64
1
Enabled
39.9 ppm
Failure
0.2785000
-0.0681700
INFO
Menu
Instruction Manual
"Calibration factors..." see Section 3-8c.
Physical measurement...
Refer to Figure 3-14, page 3-13. These
are measurements made by the analyzer
module to verify proper functioning and
appropriate flows of sample and support
gases, if any. See Section 3-7i, page 3-
18 for a complete description.
Temperature control parameters...
CLD
Temperature control
Converter set point:
Converter P gain:
Converter I gain:
Converter bias:
Converter temperature:
Detector set point:
Detector P gain:
Detector I gain:
Detector bias:
Detector temperature:
ESCAPE
MORE
HOME
Figure 3-58. Temperature Control Menu
(Screen 1)
Displays the parameters used by the
temperature control PID algorithms. Al-
lows setting of the warning alarms for the
temperatures.
Press the MORE (F3) softkey for the
Ozonator settings.
CLD
Temperature control
Ozonator set point:
Ozonator P gain:
Ozonator I gain:
Ozonator bias:
Ozonator temperature:
Minimum full-on block FET current:
ESCAPE
HOME
Figure 3-59. Temperature Control Menu
(Screen 2)
Displays the parameters used by the
temperature control PID algorithms.
Allows setting of the warning alarms for
the temperatures.
760000-C
November 2002
7.50 ppm
300 C
0.015
0.000125
0.6 C
28.6 C
51.5 C
0.100
0.000170
0.4 C
29.2 C
INFO
7.50 ppm
58.0 C
0.100
0.000170
0.4 C
29.2 C
1.000 A
INFO
Operation
3-33