Guidance And Manufacturer's Declaration - Electromagnetic Immunity - OAKWORKS PG Series User Manual

Table of Contents

Advertisement

SPECIFICATIONS

Guidance and manufacturer's declaration - electromagnetic immunity

The table is intended for use in the electromagnetic environment specified below. The customer or the
user of the table should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short inter-
ruptions and voltage
variations on power supply
input lines IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
3 A / m
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
Not applicable
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
3 A / m
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the rela-
tive humidity should be at
least 30 %.
Mains power quality should
be that of a typical com-
mercial or hospital environ-
ment.
Mains power quality should
be that of a typical com-
mercial or hospital environ-
ment.
Mains power quality should
be that of a typical com-
mercial or hospital envi-
ronment. If the user of the
table requires continued
operation during power
mains interruptions, it is
recommended that the
table be powered from an
uninterruptible power sup-
ply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical com-
mercial or hospital environ-
ment.