MAINTENANCE & SERVICE
Report Number: 100747045ATL-007
16.0
Identification, marking and documents (IEC 60601-1-2 (Ed. 3) Clause 5)
Data:
Guidance and manufacturer's declaration –electromagnetic emissions
The PROLUX TABLE is intended for use in the electromagnetic environment specified below. The customer or the
user of the PROLUX TABLE should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Report Number: 100747045ATL-007
16.0
Identification, marking and documents (IEC 60601-1-2 (Ed. 3) Clause 5)
between portable and mobile RF communications equipment and the PROLUX TABLE
The PROLUX TABLE is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the PROLUX TABLE can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the PROLUX TABLE as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
Table 1
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
EMC Report for Oakworks Inc. on the ProLux Table
19
Table 6
Compliance
Group 1
Class B
Class A
Complies

Recommended separation distances

150 kHz to 80 MHz
W
d
0,01
0,1
1
10
100
Intertek
Electromagnetic environment – guidance
The PROLUX TABLE uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
The PROLUX TABLE is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Intertek
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
=
=
1
2 .
P
d
1
0.12
0.12
0.38
0.38
1.20
1.20
3.79
3.79
12.00
12.00
Issued:
Issued:
m
800 MHz to 2,5 GHz
=
2 .
P
d
2
3 .
0.23
0.73
2.30
7.27
23.00
07/31/2012
07/31/2012
P
Page 68 of 70