Adjustment With The Display And Adjustment Module Plicscom; Short Description; Insert Display And Adjustment Module - Vega VEGADIF 65 Operating Instructions Manual

4 ... 20 ma differential pressure transmitter with metallic measuring diaphragm
Table of Contents

Advertisement

Mount/Dismount display
and adjustment module
VEGADIF 65 • 4 ... 20 mA
6 Adjustment with the display and adjustment module PLICSCOM
6
Adjustment with the display and
adjustment module PLICSCOM
6.1

Short description

The display and adjustment module is used for measured value
display, adjustment and diagnosis. It can be mounted in the following
housing versions and instruments:
All continuously measuring sensors in single as well as double
chamber housing (optionally in the electronics or connection
compartment)
External display and adjustment unit
Note:
You can find detailed information on the adjustment in the operating
instructions manual "Display and adjustment module".
6.2

Insert display and adjustment module

The display and adjustment module can be inserted into the sensor
and removed again at any time. It is not necessary to interrupt the
power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the display and adjustment module in the desired position
on the electronics (you can choose any one of four different posi-
tions - each displaced by 90°)
3. Press the display and adjustment module onto the electronics
and turn it to the right until it snaps in.
4. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The display and adjustment module is powered by the sensor, an ad-
ditional connection is not necessary.