4.4.4.1 Air Filter
The 2000IW/IW+ Integrated Welders have an input air filter which protects from
particulate matter of 5 microns or larger.
4.4.4.2 Pneumatic Connections
Air connection to the unit is made to the AIR INLET connector on the lower rear of the
Integrated Welder, with plastic pneumatic tubing. For installations using Welder/Stand
assemblies, you must provide an air filter assembly which will support to at least 100 psig
and remove particulate matter of 5 microns or larger.
4.4.4.3 Compressed Air Supply Connections
It is important to prevent impurities and moisture from entering through the air lines and
damaging your equipment, To facilitate this when installing your welder, use an air line
hookup that includes a riser and waterleg (Refer to
Connect the air line to a clean (5 micron), unlubricated, dry, compressed air supply, 35 to
100 psi (240 to 690 kPa). Use the supplied, self draining air filter with a 5 micron element.
CAUTION
Set the pressure regulator to zero before connecting or disconnecting the air supply.

Figure 4.5 Hooking Up the Air Line

48
General Warning
Permanent damage to the gauge may result if operated above the
maximum gauge reading (100 psi/690 kPa) or below the minimum
(35 psi/240 kPa).