Error codes

The HP-200 has a capability of displaying various error codes.
An error code is displayed on the auxiliary display part while blinking.

Description of error codes

Priority
Display
levels of
errors
E-11
9
E-12
10
E-13
11
E-14
12
E-15
13
E-21
4
SENSOR
ERROR
E-22
5
E-25
8
E-27
8
E-71
6
E-72
7
E-90
1
SYSTEM
E-91
2
ERROR
E-92
3
More than one error cannot be displayed at a time. If multiple errors occur, the error having the higher priority
*1:
level (smaller number) is displayed.
For E-11 to E-15, re-calibration can be performed. However, if any of E-21 to E-25 occurs during calibration,
*2:
no re-calibration can be performed.
Take remedial action in accordance with " Remedial action for error codes " (page 72).
*3:
If no temperature electrode is used (temperature sensor type is set to "non," no error is displayed.
For the relationship between errors, alarm output, FAIL signal output, cleaning output, and transmission
output, see " Output conditions for various output types " (page 63).
Tip
When an error code is displayed, the HOLD indicator blinks and transmission output becomes the
output value specified on the Setup menu. See " Set currently held value for transmission output 1 "
(page 36) and " Set currently held value for transmission output 2 " (page 37)
Error
*1
Response speed
The response of the pH electrode is low during
error
calibration using a standard solution.
Electrode
The sensitivity of the pH electrode decreases
sensitivity error
during calibration using a standard solution.
Asymmetry
Electromotive force (asymmetry potential) for pH7 is
potential error
large during calibration using a standard solution.
Standard solution
The specified standard solution is not used.
error
Standard solution
The temperature is 55°C or higher at pH10.
temperature error
Temperature
The lead wire from the temperature sensor has
sensor
electrical discontinuity.
disconnection error
Temperature
The lead wire from the temperature sensor is
sensor short-
short-circuited.
circuit error
The temperature does not fall within a range
between -5°Cand 105°C when temperature
Temperature
sensor type is set to "Auto."
measurable
The temperature does not fall within a range
range error
between -20°C and 130°C when temperature
sensor type is set individually.
Temperature
The temperature does not fall within a range
calibration error
between -20°C and 130°C.
Glass membrane
The response membrane of the pH electrode is
error
broken.
Comparison
The liquid junction of the comparison electrode is
electrode error
clogged (only applicable when the wetted pole exists).
System error
Internal communication error
The stored data such as settings and calibration
System error
values has been lost.
System error
The A/D meter malfunctions.
Description
Troubleshooting
Occurs during
*2
pH calibration
Measurement
or pH
*2*3
calibration
Temperature
calibration
Measurement
Any time
(system error)