Set Temp Display; Set Range Cut - horiba HP-200 Instruction Manual

Industrial ph meter
Hide thumbs Also See for HP-200:
Table of Contents

Advertisement

Screen 1
Screen 2
Setting
Description
range
Set calibration
"Auto",
type
"bASic"
Set calibration
1,2,3
point
"yES",
Set auto stability
"no"
"Lo",
Set stability level
"nor",
"Hi"
"yES",

Set temp display

"no"
"non",
"r1",
"r2",
Set sub-display
"r3",
"r4",
"t"
"yES",

Set range cut

"no"
Initial
Remarks
value
Select a calibration method.
Normally "Auto" should be selected.
"Auto"
For three-point calibration or
calibration using a special standard
solution, set this parameter to "bASic.
Specify a calibration point.
2
Note: This setting is only displayed when
the calibration type is "bASic."
Specify whether the automatic stability
detection is enabled.*1
"yES"
Note: This setting is only displayed when
the calibration type is "bASic.
Select an automatic stability detection
level.
For higher stability detection level,
select "Hi"; for lower one, select "Lo."
"nor"
*1)
Note: This setting is only displayed when
calibration type is set to "bASic" and
automatic stability detection to "yES.
This setting determines whether
"yES"
temperature is displayed on the auxiliary
display part.
This setting determines what is displayed
on the auxiliary display part when the
power is turned ON or when the hold
mode is changed over to the meas mode.
"t"
Note: The solution temperature is
displayed when sensor type is set to
"non" or "t."
This setting determines whether the
display is hidden when the measured
value for pH or temperature exceeds the
display range.
Display range
"yES":
"yES"
pH: 0 to 14
Temperature: 0to100(°C)
"no":
pH: −1 to 15
Temperature: −10 to 110 (°C)
(when sensor type is set to "Auto")
Temperature: −20 to 130 (°C)
(when sensor type is set to "Auto")