Quick Guide
LAQUA 200 Series
EC210/220 & PC210/220
Handheld Meters
EC210
www.horiba-laqua.com
4. Press ENT button to confirm the conductivity reading. To abort calibration,
press MEAS key.
The meter will switch to measurement
mode and the electrode icon will
appear on the screen.
5. Repeat step nos. 1 to 4 to perform calibration with the next standard solution(s).
The meter allows up to 4 points for auto calibration and up to 5 points for manual calibration.
6. Rinse the conductivity electrode with clean water and blot using lint-free
tissue to remove excess water.
7. Immerse the conductivity electrode in sample.
Make sure that the uppermost black ring in
the conductivity electrode body is immersed
in sample and there are no bubbles trapped
within the electrode.
Conductivity Calibration & Measurement
1. Rinse the conductivity
electrode with clean
water and blot using
lint-free tissue to
remove excess water.
2. Immerse the conductivity electrode in
standard solution.
Perform calibration using a
standard solution that has a
conductivity value close to the
expected sample value.
For multi-point calibration, start
with the lowest conductivity
standard. Move to standards with
increasing conductivity values.
3. Press CAL button on the meter to switch
to calibration mode.
The screen will display the calibration type briefly and the 
will start blinking until the reading stabilizes.
Press
Press

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LAQUA 200 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for horiba LAQUA 200 Series

  • Page 1 Conductivity Calibration & Measurement 1. Rinse the conductivity electrode with clean Quick Guide water and blot using lint-free tissue to LAQUA 200 Series remove excess water. EC210/220 & PC210/220 Handheld Meters 2. Immerse the conductivity electrode in standard solution. Perform calibration using a...
  • Page 2: Measurement Modes

    Cell Constant Conductivity Conductivity Measurement Modes Setting Calibration Calibration Data Auto Stable Auto Hold Real Time VIDEO VIDEO VIDEO VIDEO VIDEO VIDEO Conductivity Setup P6 CLK* • P6.2 TIME – Set time • P6.1 DATE – Set date P5 GEN •...

This manual is also suitable for:

Laqua ec210Laqua ec220Laqua pc220Laqua pc210

Table of Contents