Self Cal Menu Calibrations - HP 16534A User Reference

1-gsa/s and 2-gsa/s digitizing oscilloscope
Hide thumbs Also See for 16534A:
Table of Contents

Advertisement

Installation and Operator's Service

Self Cal menu calibrations

Self Cal menu calibrations
Load the default calibrations factors.
1
Note that once the default calibration factors are loaded, all calibrations
must be done. The calibration must be performed in the exact sequence
listed below.
N O T E
The calibration PROTECT/UNPROTECT switch on the HP 16533/34A
Oscilloscope module must be set to UNPROTECT.
Touch the Module field, then at the pop-up menu touch the Scope
a
card to be calibrated.
Touch the Menu field, then at the pop-up menu touch Calibration.
b
Touch the Mode field, then at the pop-up menu touch Service Cal.
c
Touch the Procedure field, then at the pop-up menu touch Default
d
Values.
Touch the Start field, then follow the instructions on the display.
e
Messages will be displayed as each operational accuracy calibration routine
is completed to indicate calibration has passed or failed. The resulting
calibration factors are automatically stored to nonvolatile RAM at the
conclusion of each calibration routine.