Time Base Functions
Sample Period Field
HP 16533A Sample Period Table
s/Div
Sample
Rate
5 s
100 Sa/s
2 s
250 Sa/s
1 s
500 Sa/s
500 ms
1 KSa/s
200 ms
2.5 KSa/s
100 ms
5 KSa/s
50 ms
10 KSa/s
20 ms
25 KSa/s
10 ms
50 KSa/s
5 ms
100 KSa/s
2 ms
250 KSa/s
1 ms
500 KSa/s
500 µs
1 MSa/s
200 µs
2.5 MSa/s
100 µs
5 MSa/s
50 µs
10 MSa/s
20 µs
25 MSa/s
10 µs
50 MSa/s
5 µs
100 MSa/s
2 µs
250 MSa/s
1 µs
500 MSa/s
500 ns
1 GSa/s
200 ns
1 GSa/s
100 ns
1 GSa/s
50 ns
1 GSa/s
20 ns
1 GSa/s
10 ns
1 GSa/s
5 ns
1 GSa/s
1 GSa/s
2 ns
1 ns
1 GSa/s
Notes:
Sa/s = Samples per second.
Trace length = 32768 x sample rate.
Percent on screen = The percentage of the total acquired samples that are seen on screen when the
acquired data is displayed.
4-12
Sample
Trace
Period
Length
10 ms
327.68 s
4 ms
131.07 s
2 ms
65.536 s
1 ms
32.768 s
400 µs
13.107 s
200 µs
6.553 s
100 µs
3.276 s
40 µs
1.310 s
20 µs
655.36 ms
10 µs
327.68 ms
4 µs
131.07 ms
2 µs
65.536 ms
1 µs
32.768 ms
400 ns
13.107 ms
200 ns
6.553 ms
100 ns
3.276 ms
40 ns
1.310 ms
655.36 µs
20 ns
327.68 µs
10 ns
131.07 µs
4 ns
65.536 µs
2 ns
32.768 µs
1 ns
32.768 µs
1 ns
32.768 µs
1 ns
32.768 µs
1 ns
32.768 µs
1 ns
32.768 µs
1 ns
32.768 µs
1 ns
32.768 µs
1 ns
32.768 µs
1 ns
Percent on
Screen
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
15%
6%
3%
1%
<1%
<1%
<1%
<1%
<1%

Advertisement