To Reconfigure A Multi-Card Module Into Single-Card Modules; To Turn On The System; To Test The Module; To Clean The Module - HP 16534A User Reference

1-gsa/s and 2-gsa/s digitizing oscilloscope
Hide thumbs Also See for 16534A:
Table of Contents

Advertisement

Installation and Operator's Service

To reconfigure a multi-card module into single-card modules

To reconfigure a multi-card module into single-card
modules
To reconfigure a multi-card HP 16533/34A module into single-card modules,
first remove power from the HP 16500 mainframe. Remove all
master/expander trigger cables from the rear panel of each HP 16533/34A
card in the module. Remove and reinstall each HP 16533/34A card as desired
and reapply power to the mainframe.
Performing an operational accuracy calibration on each HP 16533/34A card
is recommended.

To turn on the system

Connect the power cable to the mainframe.
1
Turn on the power switch.
2

To test the module

If you need to deskew the channel-to-channel variations, go to
"Calibrating and Adjusting", found later in this chapter.
If you require a test to initially accept the operation, perform the self
tests in chapter 3 of the optional Service Guide. The Service Guide
can be ordered through your HP Sales Office.
If you require a test to verify the specifications, start at the beginning
of the "Testing Performance" chapter in the optional Service Guide.
If the module does not operate correctly, go to the beginning of the
"Troubleshooting" found in chapter 5 of the Service Guide.

To clean the module

With the instrument turned off and unplugged, use mild soap and water to
clean the front of the module. Harsh soap might damage the water-based
paint.