Immediate Trigger Mode

When you are set up in a multi-card configuration, only the master
card input channels are recognized as the trigger source. The
oscilloscope will not trigger on inputs from the other cards in a
multi-card module.
This section discusses how immediate triggering works in a
single-card module or in the master card in a multi-card module.
Immediate trigger mode causes the oscilloscope to trigger by itself.
When you touch the Mode field, Immediate can be selected from the
pop-up menu.
Immediate trigger mode can be used for dual timebase (intermodule)
applications where, for instance, one oscilloscope module arms
another oscilloscope module, or another module (such as a logic
analyzer) arms the oscilloscope.
This mode is very similar to Auto-Trig On, but immediate mode does
not wait for a specified event to occur.