HP 16534A User Reference page 123

1-gsa/s and 2-gsa/s digitizing oscilloscope
Hide thumbs Also See for 16534A:
Table of Contents

Advertisement

Specifications and Characteristics
Supplemental Characteristics
Connect-the-dots Provides a display of the sample points connected
by straight lines.
Waveform Reconstruction When there is insufficient data to fill every
horizontal location, a post acquisition reconstruction filter fills in
missing data points for time base < 50 ns/div.
Waveform Math Display capability of A-B and A+B functions is
provided.
Measurement and Display Functions
Time Markers Two vertical markers, X and O, are provided for
measurements of time and voltage. Capabilities are: measure voltage at
point where X and O markers cross each analog waveform; measure
time from trigger to X, trigger to O, and X to O; automatic marker
placement by specifying percentage of edge, edge number, and rising or
falling edge type; run until X to O > than, < than, in range, and not in
range provides selective event search; X to O statistics (mean, max, and
min) provide analysis of time interval deviation.
Voltage Markers Two horizontal markers, A and B, are provided for
voltage measurements. These markers can each be placed on any
acquisition channel trace. A delta voltage display provides delta V
between markers on the same channel or between channels.
Automatic Search Searches for a percentage voltage level at a
positive or negative edge, count adjustable from 1 to 100.
Auto Search Statistics Mean, maximum, and minimum values for
elapsed time from X to O markers for multiple runs. Number of valid
runs and total number of runs available.
10–7