1 Application
Centrifuges are used for separation samples taken from people's, animal's and plant's components of different densi-
ties, under the influence of the centrifugal force, to provide information about their biological state (C0336 – ventilated,
C0336R– with cooling).
Its construction ensures easy operation, safe work and wide range of applications at laboratories engaged in routine
medical analyses, biochemical research works etc.
This centrifuge is not biotight and therefore during centrifugation of preparations requiring biotightness one has to
use closed and sealed containers and rotors. In the centrifuge, it is prohibited to centrifuge caustic, inflammable and
explosive preparations.
2 Technical specification
manufacturer
type
mains voltage (L1+N+PE)
mains frequency
connected load (max.)
overcurrent protection
cooling medium
capacity (max.)
speed – RPM
force – RCF
kinetic energy (max.)
running time
time counting
short-time operation mode – SHORT
continuous operation mode – HOLD
user programs
adjustable temperature
initial cooling (Fast Cool)
guaranteed temperature with max.
rotor speed
cooling/heating without centrifuging
cooling/heating with centrifuging
acceleration (ACEL)
deceleration (DECEL)
programmable non-linear curves:
acceleration
deceleration
USB communication
Electromagnetic compatibility
ambient conditions
set-up site
ambient temperature
humidity (maximum relative humidity)
excess-voltage category
pollution degree
safety area
C0336
230V
120V
10%
±5%
50/60Hz
250W
T 4A
-
00:00:01 ÷ 99:59:59 – [hours, min., sec] (step 1s)
since start button is pressed / since preselected speed is reached
-
no
-
no/no
no/no
according to EN 61326-2-6:2006
LABNET INTERNATIONAL
230V
10%
50Hz 60Hz
R507 (CFC/HCFC free)
500 ml
90 * 18000 rpm (step 1 rpm)
24270 x g (step 1 x g)
8800 Nm
yes
yes
100
-20 ÷ 40*C* (step 1°C)
10 linear curves
10 linear curves
10
10
yes
PN-EN 61010-1 p.1.4.1
indoors only
2° ÷ 40*C
< 80%
II EN 61010-1
2 EN 61010-1
300 mm
C0336R
120V
±5%
60Hz
600W
T 10A
yes
≤4*C
yes/no
yes/no
Centrifuge user manual | 1
Need help?
Do you have a question about the C0336 and is the answer not in the manual?