Selecting A Printer; Selecting The Operator Terminal; Temperature Compensation - Bosch ETT 008.31 Repair Instructions

Exhaust gas analyzer
Table of Contents

Advertisement

4.1.11 Selecting a printer

Press
"SELECT"
11 x

4.1.12 Selecting the operator terminal

Press
"SELECT"
12 x

4.1.13 Temperature compensation

Press
"SELECT"
13 x
Display
L P.Ei
Display
tE.Ei
Display
no
tc.An
7.345
7.446
7.407
i.0
tc.Ei
Meaning/function
The following settings are possible:
1 1 1 1 1 = external printer 20 characters (Switzerland)
2 2 2 2 2 = internal printer (PDR 078)
3 3 3 3 3 = external printer (PDR 200)
Meaning/function
The following settings are possible:
1 1 1 1 1 = Pocket keypad PT 14 (V24 interface)
2 2 2 2 2 = Data terminal DTL 224 (RS 232 interface)
Meaning/function
no
no" appears in the display window for engine speed, the
no no
If "no
temperature-compensation sequence has not been enabled.
To enable temperature compensation, this function must be
selected as described in Section 4.4.
Before temperature compensation, please observe the follo-
wing points:
- Let device cool for 2 hours with cover closed.
- Set up switched-off device in a calm spot free from
draughts.
- Switch on device with cover closed.
An automatic sequence takes place. The current ADC values
are shown in the display window for HC, CO and CO
i.0." appears in the engine-speed dis-
i.0.
i.0.
i.0.
approx. 100 minutes, "i.0.
play window once compensation has been successfully com
pleted.
. After
2
15