Checking And Fine-Tuning Measuring Accuracy; Taking Account Of Barometric Altitude; Checking The Indicating Accuracy - Bosch ETT 008.31 Repair Instructions

Exhaust gas analyzer
Table of Contents

Advertisement

6

Checking and fine-tuning measuring accuracy

! Before commencing testing or fine-tuning work as described in Sections 6.2 or 6.3, the analyzer must be warmed up for a period
of 10 minutes. The cover of the device must be closed during this time.
It is essential to note the instructions in Section 6.1 before checking and fine-tuning measuring accuracy.
Please observe the propane factor of 0.5 for checking or fine-tuning the HC measuring channel.
Observe safety regulations when working with toxic gases.
6.1

Taking account of barometric altitude

Before checking or fine-tuning measuring accuracy, using a barometer read the current air pressure at the site in which testing or
adjusting is taking place.
In addition, find out what altitude is stored in the analyzer in input mode to Section 4.1.7.
If the analyzer is equipped with an air-pressure gauge (and this has been activated as described in Section 4.1.6), then check the
following:
- Does the altitude stored in the analyzer correspond to a mean air pressure which does not deviate by more than +50 mbar from the
currently measured air pressure? If there is a difference of 50 mbar or more, set the altitude in the analyzer to the current value.
If the analyzer is not equipped with an air-pressure gauge, the altitude in the analyzer must be set to the value which most accurately
corresponds to the current air pressure.
After you have finished checking or fine-tuning the device, set the altitude at which the device is to be used by the operator.
Altitude for mean air pressure
Altitude in m
Mean air pressure
above sea
level
in torr
-700
825
-600
815
-500
806
-400
797
-300
787
-200
778
-100
769
0
760
100
751
200
742
300
733
400
724
500
716
600
707
700
699
800
691
900
682
1000
674
1100
666
6.2

Checking the indicating accuracy

The pump must be switched off. Select analysis mode with the "Select" button (press 10 x) and activate with the "Enter o"
button.
Introduce a flow of test gas no. 4 into the device via the calibration-gas inlet.
Read the displayed values of the CO, CO
Maximum permitted indicating error = 5% from certified test-gas value.
If the value of one of the measuring channels exceeds the maximum permitted error, re-tune the measuring channel in ques-
tion to Section 6.3.
in Hpa (mbar)
1099
1087
1074
1062
1049
1037
1025
1013
1001
989
977
966
955
943
932
921
909
899
888
and HC measuring channels and compare with the certified test-gas values.
2
Altitude in m
Mean air pressure
above sea
level
in torr
1200
658
1300
650
1400
642
1500
634
1600
627
1700
619
1800
611
1900
604
2000
596
2100
589
2200
582
2300
574
2400
567
2500
560
2600
553
2700
547
2800
539
2900
533
3000
526
in Hpa (mbar)
877
867
856
845
836
825
815
805
795
785
776
766
756
747
737
729
719
711
702
25