Ram Check Error - Bosch ETT 008.31 Repair Instructions

Exhaust gas analyzer
Table of Contents

Advertisement

8.4

RAM check error

System parameters are stored in the RAM module. During the RAM check, these system parameters are monitored to ensure that they
remain within sensible limits pre-defined by the manufacturer. If the RAM contents are outside these limits, an error message results.
A faulty RAM content may have various causes:
- A faulty powwer supply (battery)
- Replaced motherboard
- Battery replaced with the exhaust gas analyzer switched off
- Defective RAM module (IC)
- External interference signals due to inadequately shielded peripheral equipment. The connections of the temperature sensor, clamp-
on speed sensor and serial interface must be equipped with perfect shield bonding.
Check that connections comply with circuit diagram in Section 12.
The test function (see Section 4.2) allows you to determine which parameters the exhaust-gas analyzer has recognized as faulty.
Determine and correct the faulty parameters.
Press "Test u" button 7 x :
i The code numbers may appear individually or as the sum total of individual code numbers. In this way, several faulty parameters
can be displayed in a single number, e.g.: 1105 = 1 + 16 + 64 + 1024.
HC display
Faulty parameters
1
HC measuring channel mean value,
small measuring range
2
HC measuring channel mean value,
large measuring range
4
HC measuring channel upper range value,
small measuring range
8
HC measuring channel upper range value,
large measuring range
16
CO measuring channel mean value,
small measuring range
32
CO measuring channel mean value,
large measuring range
64
CO measuring channel upper range value,
small measuring range
128
CO measuring channel upper range value,
large measuring range
256
CO
measuring channel, mean value
2
512
CO
measuring channel, upper range value
2
1024
HC temperature coefficient
2048
CO temperature coefficient
O
display
Faulty parameters
2
1
Setpoint of HC calibration
2
HC calibration factor
4
Setpoint of CO calibration
8
CO calibration factor
16
Setpoint of CO
32
CO
calibration factor
2
64
Unoccupied
128
Unoccupied
256
CO
Temperature coefficient
2
512
Altitude
1024
Pump delivery range
2048
Unoccupied
Code numbers which define the faulty parameter(s) appear in the HC and O2 display fields.
calibration
2
For correction see Section:
6.3 Fine-tuning the indicating accuracy
4.1.13 Temperature compensation
For correction see Section:
For Switzerland and GB only
4.1.13 Temperature compensation
4.1.7 Entering the altitude
4.1.9 Adjusting the pump capacity