4.1
Input mode (switch S 60 at position 0, auxiliary functions)
Input mode permits you to carry out various checks and setting work on the device.
- Select the individual functions using the "Select" button. A code for each function and "An
the oil-temperature display field.
- Simultaneously press the "Enter o" button to activate the chosen function. A code for each function and "Ei Ei Ei Ei Ei " for "Input" ("Eingabe")
appears in the oil-temperature display field. After this, release both buttons.
You can now carry out modifications in the chosen function using the "Enter o" or "Test u" buttons.
- Press the "Select" button once more to return to standby mode.
Summary of select mode
i Going into the "Select modes" varies with versions for
different countries!
The settings and various fine-tuning procedures described in
these instructions are based on the standard version.
Swiss
Standard
version
version
U 13 H
U 13 U
Press
ETT 8.37,
ETT 8.31,
"SELECT" ETT 8.38
ETT 8.36
button
LE.An
LE.An
1 x
5C.An
LA.An
2 x
5C.An
02.An
3 x
5C.An
C0.An
4 x
Li.An
50.An
5 x
LA.An
Pr.An
6 x
02.An
Ho.An
7 x
C0.An
0F.An
8 x
5o.An
PL.An
9 x
Pr.An
An.An
10 x
CA.An
LP.An
11 x
Ho.An
tE.An
12 x
0F.An
tc.An
13 x
An.An
EA.An
14 x
LP.An
5t.An
15 x
tE.An
16 x
tc.An
17 x
EA.An
18 x
dA.An
19 x
5t.An
20 x
UK
French
version
version
U 13 E
U 13 F
ETT 8.31
ETT 8.35
ETT 8.52
LE.An
LE.An
5C.An
LA.An
5C.An
02.An
5C.An
C0.An
AJ.An
50.An
LA.An
Pr.An
02.An
0F.An
C0.An
Pl.An
5o.An
An.An
Pr.An
LP.An
Ho.An
tE.An
0F.An
tc.An
Pl.An
EA.An
An.An
5t.An
LP.An
tE.An
tc.An
EA.An
5t.An
An
An
An" for "Display"("Anzeige") appears in
An
Select mode functions in plain text
Display in
Function/seting mode
HC display field
LE.An
Start leak test
5C.An
Set setpoints, with automatic recognition
for test gases. Separate for HC, CO and
CO
2
Li.An
Limit value of oil temperature for official
measurements, adjustment
LA.An
Pre-select lambda calculation On / Off
C0.An
Pre-select CO
5o.An
Switch between Summer / Winter
Pr.An
Pre-select air-pressure measurement On / Off
CA.An
Pre-select calibration, weekly or each
time button is pressed
Ho.An
Set altitude
0F.An
Set oxygen zero
An.An
Analysis mode
LP.An
Select printer
tE.An
Select operator terminal
tc.An
Start temperature compensation
EA.An
External display switched on
dA.An
Select single/dual printout
5t.An
Pre-select analog output (only possible
with built-in analog interface)
PL.An
Set pump capacity
AJ.An
Start automatic calibration
02.An
Oxygen measurement switched on
calculation On / Off
vrai
9