2-Point Adjustment - E+E Elektronik EE771 Series Manual

Flow meter for compressed air and gases
Table of Contents

Advertisement

upper adjustment point at 80% of measuring range
lower adjustment point automatically at 0% of m.r.
characteristic line before adjustment
100%
characteristic line after adjustment
90%
80%
70%
60%
50%
40%
30%
20%
10%
10%
20%
30%
40%
50%
measuring range
5.4.2.

2-point adjustment

With a 2-point adjustment procedure the lower adjustment point must be between 0 and 40% of the measuring
range, and the upper adjustment point between 60 and 100% of the measuring range. If the adjustment point is
between 40 and 60% of the measuring range, automatically a 1-point adjustment procedure will be executed instead.
lower adjustment point
possibility 1
0 - <40% of m.r.
possibility 2
40 - <50% of m.r.
possibility 3
0% of m.r.
lower adjustment point at 10% of measuring range
upper adjustment point at 90% of measuring range
characteristic line before adjustment
100%
characteristic line atfer adjustment
90%
80%
70%
60%
50%
40%
30%
20%
10%
10%
20%
30%
40%
50%
measuring range
selected adjustment point
60%
70%
80%
90%
100%
upper adjustment point
60 - 100% of m.r.
100% of m.r.
50 - <60% of m.r.
60%
70%
80%
90%
100%
29
lower adjustment point at 20% of measuring range
upper adjustment point automatically at 100% of m.r.
characteristic line before adjustment
100%
characteristic line after adjustment
90%
80%
70%
60%
50%
40%
30%
20%
10%
selected adjustment point
10%
20%
30%
40%
50%
measuring range
60%
70%
80%
90%
100%